R&S®ZVA Vector Network Analyzers
Frequency ranges from 300 kHz to 50 GHz, with two or four test ports
Features
Connectivity
Various rear-panel connectors are provided to support the following:
- Control of the instrument by remote control software.
- DC and power measurements using an
R&S®NRP sensor. - Synchronization (reference frequency, external trigger, sweep and trigger status indication for direct and fast interaction with external devices).
- Sweep-time measurement: Pin 4 of the USER CONTROL port indicates the measurement duration, corresponding to the setting in the “Trigger settings” menu (sweep, sweep segment, test points, or partial measurement). By using an oscilloscope, you can easily monitor the real execution time required in each case.
| Interface | Function |
|---|---|
LAN 1 | Control of the |
LAN 2 | Control of the |
IEC-BUS | Remote control of the |
PORT BIAS | Input of DC power for supply of amplifiers via inner conductor of test port; one input and fuse per RF port; BNC connectors |
USB | Standard 1.1 USB double port (for mouse, keyboard, printer, storage media, etc.) |
10 MHz REF | Reference frequency I/O: 10 MHz reference can be input from, or output to, external devices; BNC connector |
DC MEAS | Measurement inputs for DC voltage and PAE measurements; input voltage ranges ±1 V and ±10 V; Mini DIN connectors; required cable: |
MONITOR | For external VGA monitor; standard VGA connector |
USER CONTROL | TTL I/O signals: sweep status indication, pass/fail indication, indication of active test port, indication of active channel by channel bits, trigger signals for fast control of external devices, e.g. generators, handshake with handlers, sequence control without programming effort; 5-pin D-Sub connector |